Critical Recall: Baxter Sigma Spectrum Infusion Pumps Removed Due to Software Risks

The FDA has issued a notice regarding the urgent recall of specific Baxter Sigma Spectrum Infusion Pumps due to software irregularities. This action, which aims to mitigate risks of inaccurate drug delivery, highlights serious concerns about patient safety for those using affected devices. Clinical, quality, regulatory, and healthcare teams are urged to review this update closely and take immediate action.

Affected Products and Key Details

The recall affects Baxter Sigma Spectrum V6 and V8 infusion pumps, with identified serial numbers listed below. It has been confirmed that incorrect software versions may have been installed on these devices. Spectrum V6 pumps may contain software intended for V8 devices and vice versa, leading to potential errors in infusion rates such as overinfusion or underinfusion. These deviations could significantly compromise therapy delivery, potentially resulting in drug toxicity, overdose, fluid imbalance, or suboptimal drug levels.

The following is a partial list of affected serial numbers:

  • 765100
  • 772744
  • 789742
  • 841197
  • 867753
  • 920606

Healthcare teams are advised to consult the full serial number list provided in Baxter’s communication to identify and address affected products.

Why This Recall Was Initiated

Baxter discovered discrepancies in software installation, where platforms were not paired with their correct software. As clarified in the FDA statement, this mismatch primarily results from variations between the V6 and V8 pump designs, which include distinct operational setups and clinical workflows. For example:

  • Interfaces and menu displays differ, potentially confusing users familiar with one platform.
  • Miscalibration due to software misalignment can create errors in infusion volume over extended periods.

Such risks could affect vulnerable populations exposed to inappropriate fluid or medication dosing, posing a heightened danger of serious injury or even death. Although Baxter has reported no injuries or fatalities as of June 12, proactive measures are critical given the potential severity of the issue.

Intended Purpose and Use of Spectrum Infusion Pumps

Baxter Sigma Spectrum pumps are used under medical supervision to administer intravenous fluids, blood products, and medications. These devices are integral to therapeutic delivery for patients requiring highly controlled and precise dosing.

Actions Required

Healthcare professionals must take the following steps:

  • Identify Spectrum infusion pumps with affected serial numbers and schedule their removal from service.
  • Check the software version upon powering up affected pumps. V6 software versions start with “6,” while V8 software versions start with “8.”
  • Contact Baxter’s Global Technical Services at corporate_product_complaints_round_lake@baxter.com or 800-843-7867 for further assistance on software verification and servicing.

Furthermore, institutions are encouraged to provide copies of this communication to relevant departments, including clinical leadership, pharmacy directors, and risk management teams. Contractors and OEMs distributing the affected products should notify their customers immediately.

How to Report Adverse Events

Healthcare professionals or consumers who encounter quality issues or adverse reactions are encouraged to report these incidents promptly via the FDA’s MedWatch Safety Reporting Program. Detailed reporting improves device traceability and guides ongoing regulatory review efforts to address concerns systematically.

Conclusion and Next Steps

Baxter’s recall of Sigma Spectrum Infusion Pumps underscores critical safety considerations in medical device use. Immediate removal of affected devices and verification of software accuracy are imperative steps to mitigate risks. Affected stakeholders are advised to follow FDA recommendations and work closely with Baxter to address these urgent safety concerns.

Disclaimer

This information is intended for clinical and regulatory professionals. It is not legal advice and does not replace professional consultation regarding compliance or risk management.

For full information about the FDA announcement, see the link below.
https://www.fda.gov/medical-devices/medical-device-recalls-and-early-alerts/infusion-pump-software-recall-baxter-removes-sigma-spectrum-infusion-system-platforms